Muffle Furnace -Sop for Operation and cleaning
It describes the (Standard Operating procedure) Sop for Operation and Cleaning of Muffle Furnace.
Sop for Operation and Cleaning of Muffle Furnace
I. Purpose & Scope:
- The purpose of this document is to lay down the procedure of operation, cleaning and Calibration of Muffle furnace.
- This procedure is applicable for operation, cleaning and calibration of muffle furnace used in Quality Control Department.
II. Responsibilities:
• All Microbiologist, Quality personnel shall be responsible to follow and implement this SOP.
III : Introduction and Procedural Part :
Cleaning procedure:
- Ensure that the power supply to the Muffle Furnace is switched OFF before cleaning.
- Clean the instrument with a clean dry cloth every day and occasionally wet cloth dipped in dilute soap solution may be used.
- Precautions have to be taken to clean the instrument immediately wet cloth removes the moisture.
Operation and Calibration Procedure:
Precautions :
- Ensure that the instrument is clean.
- Check the calibration status.
- Do not set the temperature above 1100°C.
- Use asbestos hand gloves and long stainless steel pair of tongs for handling.
- Use asbestos hand gloves and long stainless steel pair of tongs for handling of crucible.
- Don’t open the door frequently to avoid thermal shock.
- Initially heat the sample on hot plates in fuming chamber before keeping in muffle furnace.
- Wait until the desired required set temperature ±5°C is achieved.
Basic Operation :
- Switch ON the main supply of muffle furnace
- The red colour display will indicate the actual temperature and
- yellow colour display will indicate the earlier set temperature.
- Press the main key for changing temperature.
- For increase or decrease the temperature press arrow mark as per requirements.
- Allow the instrument to reach the set temperature.
- Open the door and keep sample crucible in the muffle furnace and follow respective procedure/specification.
- After completion of ignition as described in the procedure/ specification, the door and remove the crucible from furnace with the help of pair of tongs.
- Close the door and switch OFF the heater and main switch of the instrument.
Calibration :
- Calibration Frequency: yearly.
- Calibrate the equipment from the outside party once in a year.
- Handle the out of calibration results as per Out of specification SOP of respective unit
- Maintain the instrument usage log book as per respective SOP Handling of instrument usage log book.