Sop for Rounding of analytical results

Rounding of analytical results
Rounding of analytical results

It describes the (Standard Operating procedure)  SOP for Rounding of analytical results in Quality Control Laboratory .

Sop for Rounding of analytical results

I. Purpose & Scope:

  • To provide a general methodology to be followed for Rounding of analytical results in Quality control laboratory of pharmaceutical industry
  • Applicable to all personnel who are working in Quality control laboratory.

II. Responsibilities:

  • All Quality Control and related personnel shall be responsible to follow and implement this SOP.

III : Introduction and Procedural Part :

PROCEDURE:

  • Ensure the correctness of all entries in analytical data.
  • Ensure whether all the results are complies with the specification.
  • If the criterion meets the specification, then the observed values shall be rounded off to the number of decimal places that is in agreement with the limit expression.
  • Numbers should not be rounded until the final calculations for the reportable value have been completed.
  • Intermediate calculations (e.g. Slope for linearity) may be rounded for reporting purposes but the original value should be used for any additionally required calculation.
  • Acceptance criteria are fixed numbers and shall not be rounded off.
  • While rounding the figure, refer table no.1, 2 & 3 (Annexure-I) for decimal places to be expressed for final recording.
  • If the last digit is smaller than 5 it is eliminated and the preceding digit is Unchanged.
  • If this digit is equal to or greater than 5, it is eliminated and the preceding digit is increased by 1.
  • Consider the digit next to the intended digit for particular expression.

                                               Table No. : 01 

For Assay & Uniformity of content or uniformity of dosage units 97.3471% considers 97.3 %

97.3542 % considers 97.4 %

97.3995 % considers 97.4 %

97.8235 % considers 97.8 %

For Dissolution 97.3476 % considers 97 %

97.8238 % considers 98 %

97.5322 % considers 98 %

For Water content & LOD 0.245% considers 0.25%

0.243% considers 0.24%

1.25% considers 1.3%

For Related substances 0.105% considers 0.11%

0.056% considers 0.06%

0.0145% considers 0.01%

  • For Assay and Uniformity of Content use 1 digit after decimal point. Refer Table No.01.
  • Dissolution values shall reported complete round off. No decimal shall be used. Refer Table No. 01.
  • For Water content, LOD & Related substances use 2 digits after decimal place. E.g. Impurity is 0.145% considers 5 as rounding & value will be 0.15%.
  • For uniformity of weight report the values in 2 digits after decimal point.
  • If the specification limit for water content, LOD is ≥0 % then report the values in 2 digits after decimal point.
  • If the specification limit for water content, LOD is above the 1.0% then report the values in single digit after decimal point.
  • If the specification limit for impurities is in 2 digits after decimals point e.g. not more than 0.10% then reports the values in 2 digits after decimals point.
  • If the specification limit for impurities is in 1 digit after decimal point e.g. not more than 0.5 % then reports the values in 2 digits after decimal point and if above 1.0 % then reports the values in 1 digit after decimal point.
  • Refer to Annexure – I, which illustrate rounding numerical values for comparison.
  • For rounding the test results of assay and uniformity of content refer table no.1.0, for Dissolution refer table no.2.0 and for LOD, Related substances refer table no.3.0.
  • Recording shall be performed only once preferably in results.The rounded values shall be report on COA.

Annexure :

Annexure -I : ILLUSTRATION OF ROUNDING NUMERICAL VALUES FOR COMPARISON WITH REQUIREMENTS.

 

Annexure -I

ILLUSTRATION OF ROUNDING NUMERICAL VALUES FOR COMPARISON WITH REQUIREMENTS.

 

TABLE No. 1.0: Assay & Uniformity of Content

 

Compendial Requirement Unrounded

values

Rounded

values

Conforms
 

Assay & uniformity of content limit ≥98.0%

 

 

 

97.96%

97.92%

97.95%

98.0%

97.9%

98.0%

Yes

No

Yes

 

 

Assay & uniformity of content limit ≤101.5%

101.55%

101.46%

101.45%

101.6 %

101.5%

101.5%

No

Yes

Yes

 

TABLE No 2.0: Dissolution

 

Compendial Requirement Unrounded

values

Rounded

values

Conforms
 

Dissolution: Not less than 75%

 

74.96%

74.94%

74.95%

75.0%

74.9%

75.0%

Yes

No

Yes

 

TABLE No.: 3.0: Related substances

Compendial Requirement Unrounded

values

Rounded

values

Conforms
Related substances

Any impurity: Not more than 0.10%

 

 

 

0.095%

0.101%

0.105%

0.10%

0.10%

0.11%

Yes

Yes

No

Total impurity: Not more than 1.0% 0.96%

1.04%

1.05%

1.0%

1.0%

1.1%

Yes

Yes

No

 

 

You may also like...

Leave a Reply

Your email address will not be published. Required fields are marked *