SOP for Integrity testing Of HEPA filters
This Document describes (Standard Operating procedure) SOP for Integrity testing Of HEPA filters
SOP for Integrity testing Of HEPA filters
I. Purpose & Scope:
- The Purpose of this SOP is to lay down the procedure for Integrity testing Of HEPA filters.
- This procedure is applicable to all Engineering Department for Integrity testing Of HEPA filters.
II. Responsibilities:
- All Engineering personnel’s shall be responsible to follow and implement this SOP.
III : Introduction and Procedural Part :
Procedure Integrity testing Of HEPA filters :
Terminal HEPA Filter (having individual upstream port) :
- Start AHU of which Integrity testing Of HEPA filters checking is to be verified.
- Ensure the power supply of photometer is’ ON’
- Start the compressed air or N2 gas to generate the test aerosol at minimum pressure 20 psi.
- Direct the test aerosol at the return air point of fresh air intake of the AHU.
- Put the photometer selector switch ‘ON’ upstream mode.
- Connect the tube of photometer to the stream port of HEPA housing.
- If the upstream port is not provided then check upstream concentration at main supply duct of AHU.
- Wait until the photometer reading crosses the 22% reading then set this reading as 100% concentration, using the internal calibration button.
- Put the photometer selector switch on downstream mode.
- Wait until photometer displays ‘0’ (zero).
- Measure the downstream concentration by holding the probe approximately one inch away from face of the filter.
- Scan the corners of HEPA filter for checking the installation leakages.
- If the leakage is more than 0.01% of upstream aerosol concentration, rectify the leakage by tightening the locking bolts.
- Scan the entire filter face including perimeters with the probe of photometer in overlapping strokes, traversing at approximately 10 feet per minute ( FPM ).
- Replace the filter and check the integrity of filter as per above points
LAF Work Station / HEPA Module :
- Start the LAF of which Integrity testing Of HEPA filters is to be checked.
- Record the magnehelic gauge reading.
- Ensure the power supply of photometer is’ ON’.
- Start the compressed air to generate the test aerosol at minimum pressure 20 psi.
- Direct the test aerosol at the return air point of fresh air intake of the AHU.
- Put the photometer selector switch ‘ON’ upstream mode.
- Wait until the photometer reading crosses the 22% reading then set this reading as 100% concentration, using the internal calibration button.
- Put the photometer selector switch on downstream mode.
- Wait until photometer displays ‘0’ (zero).
- Measure the downstream concentration by holding the probe approximately one inch away from face of the filter.
- Scan the corners of HEPA filter for checking the installation leakages.
- If the leakage is more than 0.01% of upstream aerosol concentration, rectify the leakage by tightening the locking bolts.
- Scan the entire filter face including perimeters with the probe of photometer in overlapping strokes ,traversing at approximately 10 feet per minute ( FPM ).
- Replace the filter and check the integrity of filter as per above points.
AHU / Terminal Mounted HEPA Filter :
- Start AHU of which Integrity testing Of HEPA filters checking is to be checked.
- Ensure the power supply of photometer is’ ON’.
- Start the compressed air to generate the test aerosol at minimum pressure 20 psi.
- Direct the test aerosol at the return air point of fresh air intake of the AHU.
- Put the photometer selector switch ‘ON’ upstream mode.
- Check the upstream concentration of PAO before HEPA filters at AHU/Plenum Wait until the photometer displays 100% upstream concentration.
- Remove the tube of photometer and seal the AHU/Plenum port and ensure to zero leakage through port.
- Before entering AHU/Plenum he / her should wear apron, mask.
- Put the photometer selector switch on downstream mode.
- Wait until photometer displays ‘0’ (zero).
- Open the AHU/Plenum door and enter inside.
- Measure the downstream concentration by holding the probe approximately one inch away from face of the filter.
- Scan the corners of HEPA filter for checking the installation leakages.
- If the leakage is more than 0.01% of upstream aerosol concentration, rectify the leakage by tightening the locking bolts.
- Scan the entire filter face including perimeters with the probe of photometer in Overlapping strokes, traversing at approximately 10 feet per minute (FPM).
- Replace the filter and check the integrity of filter as per above points
Frequency for Integrity testing Of HEPA filters-
HVAC system– Yearly ± 30 days and LAF system- Six Monthly ± 15 days.