SOP for Integrity testing Of HEPA filters

This Document describes (Standard Operating procedure) SOP for Integrity testing Of HEPA filters

SOP for Integrity testing Of HEPA filters

I. Purpose & Scope:

  • The Purpose of this SOP is to lay down the procedure for Integrity testing Of HEPA filters.
  • This procedure is applicable to all Engineering Department for Integrity testing Of HEPA filters.

II. Responsibilities:

  • All Engineering personnel’s shall be responsible to follow and implement this SOP.

III : Introduction and Procedural Part :

Procedure Integrity testing Of HEPA filters :

Terminal HEPA Filter (having individual upstream port) :

  • Start AHU of which Integrity testing Of HEPA filters checking is to be verified.
  • Ensure the power supply of photometer is’ ON’
  • Start the compressed air or N2 gas to generate the test aerosol at minimum pressure 20 psi.
  • Direct the test aerosol at the return air point of fresh air intake of the AHU.
  • Put the photometer selector switch ‘ON’ upstream mode.
  • Connect the tube of photometer to the stream port of HEPA housing.
  • If the upstream port is not provided then check upstream concentration at main supply duct of AHU.
  • Wait until the photometer reading crosses the 22% reading then set this reading as 100% concentration, using the internal calibration button.
  • Put the photometer selector switch on downstream mode.
  • Wait until photometer displays ‘0’ (zero).
  • Measure the downstream concentration by holding the probe approximately one inch away from face of the filter.
  • Scan the corners of HEPA filter for checking the installation leakages.
  • If the leakage is more than 0.01% of upstream aerosol concentration,   rectify the leakage by tightening the locking bolts.
  • Scan the entire filter face including perimeters with the probe of photometer in overlapping strokes, traversing at approximately 10 feet per minute ( FPM ).
  • Replace the filter and check the integrity of filter as per above points

LAF Work Station / HEPA Module :

  • Start the LAF of which Integrity testing Of HEPA filters is to be checked.
  • Record the magnehelic gauge reading.
  • Ensure the power supply of photometer is’ ON’.
  • Start the compressed air to generate the test aerosol at minimum pressure 20 psi.
  • Direct the test aerosol at the return air point of fresh air intake of the AHU.
  • Put the photometer selector switch ‘ON’ upstream mode.
  • Wait until the photometer reading crosses the 22% reading then set this reading as 100% concentration, using the internal calibration button.
  • Put the photometer selector switch on downstream mode.
  • Wait until photometer displays ‘0’ (zero).
  • Measure the downstream concentration by holding the probe approximately one inch away from face of the filter.
  • Scan the corners of HEPA filter for checking the installation leakages.
  • If the leakage is more than 0.01% of upstream aerosol concentration, rectify the leakage by tightening the locking bolts.
  • Scan the entire filter face including perimeters with the probe of photometer in overlapping strokes ,traversing at approximately 10 feet per minute ( FPM ).
  • Replace the filter and check the integrity of filter as per above points.

AHU / Terminal Mounted HEPA Filter :

  • Start AHU of which Integrity testing Of HEPA filters checking is to be checked.
  • Ensure the power supply of photometer is’ ON’.
  • Start the compressed air to generate the test aerosol at minimum pressure 20 psi.
  • Direct the test aerosol at the return air point of fresh air intake of the AHU.
  • Put the photometer selector switch ‘ON’ upstream mode.
  • Check the upstream concentration of PAO before HEPA filters at AHU/Plenum Wait until the photometer displays 100% upstream concentration.
  • Remove the tube of photometer and seal the AHU/Plenum port and ensure to zero leakage through port.
  • Before entering AHU/Plenum he / her should wear apron, mask.
  • Put the photometer selector switch on downstream mode.
  • Wait until photometer displays ‘0’ (zero).
  • Open the AHU/Plenum door and enter inside.
  • Measure the downstream concentration by holding the probe approximately one inch away from face of the filter.
  • Scan the corners of HEPA filter for checking the installation leakages.
  • If the leakage is more than 0.01% of upstream aerosol concentration, rectify the leakage by tightening the locking bolts.
  • Scan the entire filter face including perimeters with the probe of photometer in Overlapping strokes, traversing at approximately 10 feet per minute (FPM).
  • Replace the filter and check the integrity of filter as per above points

Frequency for Integrity testing Of HEPA filters-

HVAC system– Yearly ± 30 days and LAF system- Six Monthly ± 15 days.

 

 

You may also like...

Leave a Reply

Your email address will not be published. Required fields are marked *