Conductivity Meter-SOP for Operation and Calibration
This Document describes (Standard Operating procedure) Sop for Operation and Calibration of Conductivity meter
Sop for Operation, Cleaning and Calibration of Conductivity meter
I. Purpose & Scope:
- The Purpose of this SOP is to lay down the procedure to lay down a procedure for Operation, Cleaning & Calibration of Conductivity meter.
- This procedure is applicable for Operation, cleaning & Calibration of Conductivity meter used in Quality Control Department
II. Responsibilities:
- All Quality Control personnel’s shall be responsible to follow and implement this SOP.
III : Introduction and Procedural Part :
IV) GENERAL CLEANING :
- Clean the instrument with a clean dry cloth every day. Occasionally wet cloth dipped in dilute soap solution may be used.
- Precautions has to be taken to clean the instrument immediately with dry cloth to remove the moisture.
V) BASIC OPERATION OF CONDUCTIVITY METER :
- Ensure the electrode connected to the instrument.
- Switch on the main power supply to Instrument and then on the power switch provided on the rear side of the instrument.
- Remove the electrode from purified water beaker, Clean the electrode with tissue paper.
- Dip the electrode carefully in sample beaker / solution in a center position.
- By Pressing the Range key, set the mode as Conductivity ( µs) and Conductivity value will be appear on
- Record the observed Conductivity value on analytical record or as per requirements.
- Remove the electrode from sample & clean with purified water properly and then finally clean with tissue paper.
- Keep the electrode in stand mode in water beaker containing purified water.
VI) CALIBRATION OF CONDUCTIVITY METER :
- Operate the instrument as per above basic operation procedure
- Preparation of calibration standard solution 0.1 M KCl Solution (A): Weigh 0.7456 gm of Potassium chloride and dilute to 100 ml with purified water.
- Preparation of calibration standard solution 0.01 M KCl Solution (B): Weigh 0.07456 gm of Potassium chloride and dilute to 100 ml with purified water.
- Prepare the calibration standard solutions on daily basis before use for calibration.
- The Potassium chloride shall be kept in desiccators over silica. Label the standard solution with name, strength, prepared by, prepared on and use before. Discard the solution after calibration.
- Determine the conductivity of each solution at 25°C for calibration.
- The standard solution preparation and its conductivity at 25°C is given as.
Sr. No. | Standard Solution | Acceptance Criteria + 1 % of standard value |
1 | 0.1 M KCL Solution (A) | 12.88 ± 1% ms (12.75 to 13.00 ms) |
2 | 0.01 M KCL Solution (B) | 1413 ± 1% µS (1399 to 1427 µS) |
- Remove the electrode form purified water beaker; clean the electrode with clean tissue paper.
- Dip the electrode carefully in Standard solution A (0.1 M KCL) in a center position.
- By Pressing the Range key, set the mode as Conductivity (µs) and Conductivity value will be appear on display.
- Record the observed Conductivity value Annexure I
- Remove the electrode from Standard solution A and clean with purified water properly and then finally clean with tissue paper.
- Dip the electrode carefully in Standard solution B (0.01 M KCL) in a center position.
- By Pressing the Range key, set the mode as Conductivity (µs) and Conductivity value will be appear on display.
- Record the observed Conductivity value Annexure I
- Remove the electrode from Standard solution B and clean with purified water properly and then finally clean with tissue paper.
- Keep the electrode stand in purified water.
- If any discrepancy observed during operating the instrument report to QC Manager & notify the defect to Maintenance Manager to rectify the defect. Affix ‘Under Maintenance Label’ on the instrument.
Calibration Frequency:
- Daily morning before use
VII) ANNEXURE:
Annexure I : Daily Instrument Calibration Record –Conductivity Meter.
Annexure II : Preparation record for calibration standards
ANNEXURE-I
DAILY INSTRUMENT CALIBRATION RECORD – CONDUCTIVITY METER
Name of Instrument : Conductivity Meter | Month : |
Make : | Model : |
Department: Quality Control | Instrument ID No : |
Frequency : Daily |
Acceptance Criteria:
1) For Standard Solution of 0.1 M KCL: 12.88 ± 1% mS (Limit: 12.75 to 13.00)
2) For Standard Solution of 0.01 M KCL: 1413 ± 1% πS (Limit: 1399 to 1427)
Date | Observed Conductivity | Remark | Done By | Checked By | |
For 0.1 M KCL | For 0.01 M KCL | ||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply |
Date | Observed Conductivity | Remark | Done By | Checked By | |
For 0.1 M KCL | For 0.01 M KCL | ||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply | |||||
Complies/Does not Comply |
ANNEXURE-II
PREPARATION RECORDS FOR CALIBRATION STANDARDS
Month :
Date | Weight Taken | Volume prepared
(ml) |
Prepared By | Checked By | |
For 0.1 M KCL | For 0.01 M KCL | ||||