Sop for Operation and Calibration of FTIR

Calibration of FTIR
Calibration of FTIR

This Document describes (Standard Operating procedure) Sop for Operation and Calibration of FTIR

Sop for Operation and Calibration of FTIR

I. Purpose & Scope:

  • The Purpose of this SOP is to lay down the procedure to lay down a procedure for Operation, Cleaning & Calibration of FTIR
  • This procedure is applicable for Operation, cleaning & Calibration of FTIR used in Quality Control Department

II. Responsibilities:

• All Quality Control personnel’s shall be responsible to follow and implement this SOP.

III : Introduction and Procedural Part :

GENERAL CLEANING :

  • Clean the instrument with a clean dry cloth every day. Occasionally wet cloth dipped in dilute soap solution or IPA may be used for cleaning of outer surface of instrument.
  • A precaution has to be taken to clean the instrument immediately with dry cloth to remove the moisture.

PRECAUTIONS :

  • Do not touch to clean any optical surface while cleaning.
  • Do not stare into the internal laser beam under the instrument cover.
  • Do not use a flammable gas to purge the instrument.
  • Do keep the spectrum one dry and avoid spilling liquid into the instrument.
  • Do not keep any traces of materials on the sample port of Instrument, as excess humidity may damage sensitive optics.

BASIC OPERATION OF FTIR :

  • Ensure that the power supply to the instrument is ON and Select the Peak Miracle IR ‘Instrument setup.
  • Select Software icon ‘Horizon MB’ & Open, select the Instrument configuration as MAC-00-02-2C-001CB2.
  • Fill the sample information for background (Blank) first, with comment as background.
  • Select the Instruments Parameters
  • Resolution = 4
  • Scan          = 16
  • Click on acquisition mode as Single Beam
  • Click on icon ‘Reference’ Instrument scanning is started for Blank.
  • After the completion or Back ground, Start for Sample set up & preparation.
  • Put sample/ Standard on zinc mirror to the Peak Miracle, Take care the Zinc mirror is covered by the sample.
  • Fill the sample/ Standard information for with name of sample with batch details or Analytical report number & comment as the test name.
  • Select the Instruments Parameters
  • Resolution = 4 and Scan = 16
  • Click on acquisitions mode as Transmittance.
  • Click on icon ‘Measure’ Instrument scanning is started for Sample & saves with given name.
  • Select icon for Print ‘Window –Project- Measure Data – Data, Click on the sample/ Standard file, open it & take print for the same Take print & compare the test sample & standard sample.
  • Clean the sample port with Isopropyl alcohol (IPA) after the use of instrument.

CALIBRATION OF FTIR :

Calibration of FTIR Frequency: Once in months and after each maintenance job

Verification of the wave-number Scale:
  • Select the sample compartments for polystyrene film.
  • Take Blank is air and take scan for 4000 – 400 cm-1.
  • Insert the polystyrene film in sample compartments and go to beam and select polystyrene for scan with following set up.
  • Go to scan and give the following parameters
  • Range                          –           4000 – 400 cm-1
  • Scan type                    –           Background/Sample
  • Units                           –           % T
  • Scan No.                     –           1
  • Click to scan
  • The peak data of the spectrum of the polystyrene film is displayed on the screen.
  • Print all the data from peak tables from ‘view’. And check limits of wave number, Accuracy, Compare the wave number accuracy with following limit.
  • 3060.0 ± 1.0 cm-1
  • 2849.5 ± 1.0 cm-1
  • 1942.9 ± 1.0 cm-1
  • 1601.2 ± 1.0 cm-1
  • 1583.0 ± 1.0 cm-1
  • 1154.5 ± 1.0 cm-1
  • 1028.3 ± 1.0 cm-1
  • Record the observations as per Annexure I during calibration of FTIR.
Control of resolution performance :
  • Select the sample compartments for polystyrene film.
  • Take Blank is air and take scan for 2900 cm-1 to 2800 cm-1 &  1600 cm-1 to  1500 cm-1 separately one by one.
  • Insert the polystyrene film in sample compartments and go to beam and select polystyrene for scan with following set up during calibration of FTIR.
  • Go to scan and give the following parameters
  • Range               –   2900 cm-1 to 2800 cm-1 & 1600 cm-1 to 1500 cm-1
  • Scan type        –           Background/Sample
  • Units               –           % T
  • Scan No.         –           1
  • Click to SCAN
  • The peak data of the spectrum of the polystyrene film is displayed on the screen.
  • Print all the data from peak tables from ‘view’. And check limits of wave number.
  • The difference between the absorbance at the absorption minimum at 2870 cm-1 and the absorption maximum at 2849.5 cm-1 is greater than 0.33. The difference between the absorbance at the absorption  minimum at 1589 cm-1 and the absorption maximum at 1583 cm-1 is greater than  08.
  • Record all the calibration data in Annexure-I during calibration of FTIR
  • Handle the out of calibration results as per respective SOP
  • If any discrepancy observed during operating the instrument report to QC Manager & notify the defect to Maintenance Manager to rectify the defect. Affix ‘Under Maintenance Label’ on the instrument.
  • Maintain the instrument usage log book as per SOP Handling of instrument usage log book

ANNEXURE:

Annexure I      :  Instrument Calibration Record

 

ANNEXURE-I

  Calibration of FTIR Record

 

   Name of Instrument : FTIR Make :
   Instrument ID No. Model :
Calibration done on: Calibration Due on:

 

  1. A) Cleanliness of the Instrument:

Observation: Clean / Not Clean

  1. Any other maintenance:

Observation: Required / Not required

  1. B) VERIFICATION OF WAVE NUMBER SCALE:
Sr.

No.

Claimed wave Number Observed wave Number Acceptance Criteria

(± 1.0 cm-1)

1 3060.0 cm-1   3059.0 cm-1 to 3061.0 cm-1
2 2849.5 cm-1   2848.5 cm-1to 2850.5 cm-1
3 1942.9 cm-1   1941.9 cm-1to 1942.9 cm-1
4 1601.2 cm-1   1600.2 cm-1 to 1602.2 cm-1
5 1583.0 cm-1   1582.0 cm-1 to 1584.0 cm-1
6 1154.5 cm-1   1153.5 cm-1 to 1155.5 cm-1
7 1028.3 cm-1   1027.3 cm-1 to 1029.3 cm-1

 

  1. CONTROL OF RESOLUTION PERFORMANCE:
Sr.

No.

Claimed wave No. Result Acceptance Criteria
1 The differences between the absorbance at  the absorption minimum at 2870 cm-1 and the absorption maximum at  the 2849.5 cm-1   Greater than 0.33
2 The differences between the absorbance at  the absorption minimum at 1589 cm-1 and the absorption maximum at  the 1583 cm-1   Greater than 0.08

 

Remark: The calibration of FTIR performed as per SOP and Complies / does not comply with the acceptance criteria.

Calibrated By:                                      Checked By:                               Approved By:

 Date                                                       Date:                                            Date:

 

You may also like...

Leave a Reply

Your email address will not be published. Required fields are marked *